Abstract

Dental implants provide a continuous interface between the oral environment and the deep core structures of the jaws. Implants and trans-mucosal superstructures are primarily metal and heat conduction occurs readily. A hypothetical heat conduction model is investigated to determine the ranges of temperature gradients that might occur in implants. This model showed that a 60°C heat source will cause a heat front of ⩾47°C to advance >3 mm down an implant within one second. Oral temperature transients may be a factor in implant pathology.

Keywords
Dental implants; Oral temperature; Heat conduction; Bone pathology